Replacing the lung bladders
Replace the lung bladders if they leak or are damaged.
Before starting, follow the Open the torso skin procedure.
Lift the chest and chest rise plates.
Unhook the green lung compliance bands on each side of the lung assembly.
Lift the hinged lung plate.
Remove the used lung from the socket.
Discard the used lung.
Select the new lung bladder.
Insert the new lung bladder into the socket.
Ensure that the string is placed in the lung bladder fold.
Close the hinged lung plate.
Secure the green compliance bands.